Setup Procedure for X12B Motorized CrystalLogic Goniometer Head |
| Purpose | The motorized goniometer head at X12B uses three jacks to position the crystal mounting platform in height and angle. When a crystal cap and pin are in place these motions translate into an x, y, and z crystal adjustment of sufficient range for automounter applications. Each jack may travel from 2mm to 10mm. The 5mm extension is the home position. Because the micro motors in this goniometer head have no limit switches or encoders, it is necessary to check its setup when putting the head into service. |
For more info see the Gonio Tech Pages
|
| Procedure |
|
| Goniometer Epics Medm |
![]() Gonio Jack 1, 2, and 3 are physical motors, Gonio X, Y, and Z are virtual motors. |
![]() |
|
| Updated 4 Oct 2005 |