Nov 2009
Flux measurements at X25 and X29.
These measurements were done with a photo-diode.
Tables

|
|
We varied the
apertures (slits) and measured the flux for a wide range of wavelength. Be aware for now, if you
want to screen at longer lambda than 1.2A,
it's fine but if you want to collect you should let us know in order to
optimize the intensity and the shape of the beam.
Normal mode
is 50x50 um2--brown line and the flux is 2-3x1011
photon/sec
The dashed lines is to
show the normalized to 100x100 um2
The native size of the beam (if we didn't have slits in) would be FWHM
~ 90 um vertical and ~260 um horizontal.
|

|
|
Comparison
with X29
On this graph you can see the normal mode of operation for X25 (red) and the normal mode of
operation for X29 (blue dash).
At this point "normal mode" means different things for different
beamline.
slits
size 50x50 um2
vs 160x100um2
divergence
of the beam
0.3-0.5 vs 0.4
different beam characteristics
90-260 um2
vs 130x260 um2
The idea is that users can control most of these parameters and
customize them to fit their need.
ex. easy defocus procedure
|
|
|
|
26-27 August 2009 Horizontal Divergence Measurements
Scattering setup: The beamline energy
was calibrated at the Se edge. The pass energy of the four bounce
divergence analyzer (880) reflection was found at 13415 eV. The unit
was set to scatter in the horizontal.
mono focus = 2.51
beam spot at sample (as measured with
table scans)
344 um FWHM horz (dromedary) 85 um FWHM vert (dromedary)
FEA 3000 um horizontal
colli slits 20 um horz x
50
um vert
E=13.407- 13.421 kev
mirror bend cyl. =170,000steps |
 |
The
following graphs show the
divergence versus energy.
Due to the strong sagittal focusing, a
significant correlation between angle and energy is observed.
Therefore, divergence scans were taking at energy intervals 2 eV
apart through the energy width of the monochromator bandpass.
The
FWHM of the divergence envelope is ~317 urad. The right graph shows
the integrated intensities at each energy, which gives the energy
resolution; it is observed to be ~ 9 eV FWHM.
These measurements are consistent
|
colli
slits 100 um horz x 100
um vert
E=13.407- 13.421 kev
mirror bend cyl. =170,000steps |
 |
Same measurements done on
June 2009
Divergence for focus at 2.51 ~380 urad,
the horizontal slits in the hutch are different.
|
|
|
|
Then the sagital focus was brought to -0.4 in an attempt to get the
second xtal flat. The data shows that there
is only a very slight correlation between the horizontal divergence
and the energy when the monochromator is defocused. The FWHM of the
horizontal divergence is 42.3 urad.
Keeping the
horizontal position the same, the effect of varying the vertical
mirror focus was investigated. The vertical mirror bend was changed
from 170,000 to 210,000 and the scans repeated. This value gave a
somewhat better vertical focus. The results are identical. . A
slight correlation between angle and energy is observed, and the
divergence FWHM is 45.7 urad. To check this, the mirror focus was
then changed back to 170,000, and an identical divergence scan was
measured.
23-24 June 2009
Horizontal divergence of the beam measured with a four bounce crystal
set in a dispersive geometry by Joe Dvorak.
The divergence
measurements were made with a four bounce crystal arrangement set in
a dispersive geometry. As such, the unit acts as an extremely narrow
energy and angle filter. By rotating the entire unit about the
center of the first crystal, the divergence of the beam can be
measured directly. The rotational scan stage has a resolution of
0.33 urad/step, more than adequate for this beamline. The unit was
set up to scan in the horizontal plane. The crystals are Si(220),
incident angle on the crystals is 75o
to grazing, and the included angle between the dispersive crystals is
30o.
|
|
|
|
Collimator slits = 100/100 um
White beam vertical
Front End aperture slit settings: 1.38 mm
Photon energy = 10.057 eV near Zn
|
 |
The
following graph shows
the dependence of the horizontal divergence on the white beam
horizontal slit settings (FEA HORIZONTAL).
The main graph shows the experimental
data. The inset shows how the FWHM varies with the horizontal
aperture. The results are reasonably linear over the range measured.
|
Front
End Aperture Horizontal
FEA=1.9mm
|
 |
Next, we kept the white
beam aperture fixed and varied the horizontal focus of the
monochromator to measure the effect that the focus position had on
the beam divergence.
The measurements were done at several FEA HORIZONTAL settings:
1.9, 2.9, and 4.9 mm.
The following graphs
show the experimental data, with the insets showing the variation of
the FWHM with the monochromator focus. |
Front
End Aperture
FEA=2.9mm
Normal Mode of Operation
|
 |
|
FEA=4.9mm |
 |
The
results show that when
the beam is defocused (decreasing focus setting below the nominal
“best” focus setting of 1.02 mm for this wavenlength), the divergence
decreases
approximately linearly with the focus setting, and can easily be
decreased by a factor of two or more.
This all comes at a cost in
flux, of course. It was also observed that the flux (as measured by
the “merit”) is directly proportional to the divergence. As the
white beam horizontal aperture is increased, the FWHM of the
decreases with defocusing as expected, however, the absolute value of
the divergence maximum begins to vary with the focus setting.
That
is, the peak is always found within the expected envelope, but the
exact position within that envelope becomes variable. It appears
that the sagittal focusing crystal is very sensitive to any
perturbations. |
|
|
|
June18 2009.
After the May shutdown, X25 started with beam in the hutch despite huge
work done on the mirror bender. Unfortunately the vertical beam looked
doubled.
The paddle, right after the mono looked like it didn't have enough
motions anymore to show
the beam. It needed to go higher.
Shu, Lonny, opco , Kim Hayes, Marc and I (Annie) were there because we
needed
to unlock the padlock and have radiation people present while moving
the paddle. Ultimately we didn't have to physically move the padlock
and the beam was visible and looked like 3 orange stripes. We
actually would like to see a flat fat beam. If it were to have an angle
then we would
have some issue with where the beam hits the first xtal. Not the case
here.
We are at the selenium inflection point with slits of 100x100 um2 (
merit of 489 for I=217).
When we scan the vertical beam we can see two peaks. Moving the mirror
in the vertical motion (initial position -0.88 to -0.68).
after going up by steps of 0.2 , we see the two humps seperating and
the intensity of both peaks is going down too.
Let's move down. eventually the two humps are about the same intensity.
We then turn the mirror bender (cylindrical) on and start moving it
by 10 000 steps. The temperature of the motor (look on display on top
of mirror tank) rises to about 44C when we move the bender. We end up
unbending the mirror. We get a single peak with a lot more counts.
We move the wavelength from 0.9795A to 1.1A and we still have a nice
single
peak with, according to the scan, FWHM_v says 45um. Should be at best
50-60 um.
Final bender position: 170 000, height z=-1.78.
Jan 26 2009.
Tested with a diffraction pattern what the spot shape would look if we
slit down the front end apertures.
B. Sweet--The beam-size tool I sent you on Friday predicted that the
spot should be
160um fwhm with the FE slit at 4.9mm. Indeed that's what we saw
for a
"typical" fully-recorded spot. However the spot got no smaller
when we
narrowed the slit to 1.9 and 1.3mm, where the spot is predicted to be
60-75mu. What we >did< see, however, was that the
peak:background ratio
appeared to increase from about 4 with the wide slit to 6 with the
narrower ones.
A. Heroux-These shots were not done taken with dose in
mind to compare the
intensity of the peak or the backgroung. As seen on the first row
trippling the exposure time barely the backgroung but double the
intensity of the peak.
On the second row the backgroup doubles but the paek at least triples
(if
not quadruples).
Jan 23 2008:
knife edge scans to determine the divergence of the beam at
X25 while close the front end slits. Measuremnets done by Grace and Bob.
These results are with non-calibrated front-end slits and "20 um"
apertures.
largest spacing is using lambda 0.9795A, a distance of 350mm for the
detector apertures of 100x100 um2
|
Front-end slit width/mm |
FWHM / mm |
width –
aperture |
FWHM divergence / mrad |
spot size / microns |
largest spacing |
relative intensity |
|
|
|
|
|
|
|
|
|
1.31 |
0.0254 |
0.0054 |
0.0108 |
204 |
600 |
350 |
|
1.51 |
0.0413 |
0.0213 |
0.0426 |
215 |
500 |
650 |
|
1.71 |
0.0540 |
0.0340 |
0.0680 |
224 |
500 |
840 |
|
1.91 |
0.0699 |
0.0499 |
0.0997 |
235 |
500 |
1160 |
|
2.11 |
0.0794 |
0.0594 |
0.1188 |
242 |
500 |
1360 |
|
2.71 |
0.1080 |
0.0879 |
0.1759 |
262 |
400 |
1700 |
|
3.51 |
0.1365 |
0.1165 |
0.2331 |
282 |
400 |
1900 |
|
3.91 |
0.1588 |
0.1388 |
0.2775 |
297 |
400 |
2480 |
|
4.31 |
0.1746 |
0.1546 |
0.3093 |
308 |
400 |
2600 |
|
4.91 |
0.2381 |
0.2181 |
0.4363 |
353 |
300 |
3100 |
|
|
λ / a = angular spot separation |
|
|
|
|
|
|
|
spot size = aperture + div.*dist. |
|
|
|
|
|
|
|
a = λ / (size/distance) * factor = λ / factor *
[(aperture + div.*dist.)/dist.] |
|
|
|
|
|
|
|
|
|
|
|
|
|
Summary
of Tests performed on X25 during the week of 9/29-10/3/08:1 done by
Grace)
Started in
2.584Gev Ops. The mono lift , mono height, mirror height, twists, roll
and focus were realigned to optimized the beam shape on the mono and
mirror phosphors.2) The mono lift and mono height adjustments
successfully eliminated the double peak previously observed during
pitch scans.3) The diffractometer was realigned to optimize counts. The
diffyaw and diffpitch were optimized. The FrontEnd slits were closed up
to decrease scattering.4) After the Energy was calibrated, the proper
gap was measured. Chris then calculated the needed offset and set this
in software.5) Preliminary settings for Roll and Focus were looked also
looked at just prior to the ring energy switching back from 2.584Gev to
2.8Gev ops. 6) The mirror focus was also looked at. Although, the beam
can be overfocused, attempts to underfocus the beam had unexpected
results. Additionally the elliptical bend motion continues to give
false hard limit trips. Long term our recommendation is to use the
BC-005 (APS transition board) in place of Bill's IO58 box (in the white
electronic rack), as it's wiring scheme has been more reliable. 7) The
alignment was rechecked, the Energy was recalibrated, and the gap
offset was optimized for 2.8Gev ops 8) The focus table was established
for energies ranging from 6.0 to 15 kev, and the roll motion was
further optimized.9) Bob and John L. made modifications to the
periscope to accommodate a protective lens cover. This required the
periscope motions and focal distances to be rechecked.10) A burn was
taken on the Hutch Ion Chamber to confirm the beams position and the
beam position monitor and attenuator were reinstalled. 11) On Friday
Stu cooled down the detector and the cold stream was put back in place.
12) Joe D., Annie, and Chris worked on measuring the beam flux using
Lonny's standard ion chamber. As the focus table was not fully
implemented, hand tweaking was needed for these tests.13) Chris fitted
the focus data to determine the best curve to calculate the focus for
energy moves. Although this was not completely implemented until Friday
evening, the focus table did appear to be working repeatably for .979 A
and 1.1A. 14) As mentioned in my note at the beamline, the roll will
still need some final tuning and the focus table will need additional
testing. 15) If time allows, I would recommend completing the
roll/focus testing with the existing slit calibration to maintain a
consistent merit comparison..16) As the slits have had excessive use,
the slit calibration will need to be rechecked prior to user
operations. If new offsets are used please record these in the beamline
logbook.
Summary of Sagital Focus
values and expected Merits for changing Energy on X25:
|
|
|
|
|
|
|
|
(10/6/08) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Energy |
Wavelength |
Theta |
Focus |
Roll |
diffperp |
optics_V |
Colli |
Hutch |
Merit |
Ring |
H-FWHM |
V-FWHM |
diffpitch |
diffyaw |
slits |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 6.000 |
2.0664 |
11.40275 |
-0.300 |
-0.218 |
1.5215 |
-1.4300 |
128090 |
516824 |
738 |
170ma |
243um |
|
0.49 |
-0.36 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 7.113 (Fe) |
1.7430 |
16.13710 |
0.020 |
-0.218 |
1.1217 |
-1.4400 |
92620 |
509328 |
541 |
169 |
|
*265 |
0.49 |
-0.36 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 9.000 |
0.6200 |
|
0.620 |
-0.280 |
2.8520 |
-1.3400 |
139000 |
485000 |
536 |
257 |
254 |
244 |
0.35 |
-0.37 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 10.000 |
1.2399 |
11.40275 |
1.020 |
-0.218 |
0.1419 |
-1.1200 |
118901 |
188708 |
651 |
~179 |
239 |
|
0.49 |
-0.36 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 11.000 |
|
|
1.420 |
-0.218 |
1.9789 |
-1.0470 |
139000 |
238000 |
865 |
161 |
231 |
152 |
0.33 |
-0.25 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 12.658 (Se) |
0.9795 |
8.98580 |
2.150 |
-0.218 |
-0.3081 |
-1.0900 |
148501 |
141989 |
892 |
165 |
~236 |
*107 |
0.49 |
-0.36 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 13.777 |
0.9000 |
8.25126 |
2.650 |
-0.222 |
-0.3778 |
-1.1400 |
57386 |
65832 |
302 |
~186 |
258 |
|
0.49 |
-0.36 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 15.000 |
0.8266 |
7.57388 |
3.250 |
-0.218 |
-0.5783 |
-1.0600 |
75820 |
79801 |
407 |
184 |
239 |
|
0.49 |
-0.36 |
100x100 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Bolded values reflect the most recent compatible "diffperp" positions
using the same roll, diffpitch and diffyaw values.
Oct 3, 2008
After new upgrades of the motor controls for the mono.
There doesn't seem to have anymore flickering of the beam when we wlook
at the phospho in the hutch. The motors ahve all been calibrated and we
have a new algorithm for the focus.
Measurements of the flux by Joe Dvorak.
The photodiode is set at the sample position 10^6 V/amp, 4000 cv bias.
raw data
|
C4 counts
|
Current
uA
from photodiode
|
flux
(normalize at 235mA)
photons/sec
|
ring
current mA
|
Flux
(to come)
|
0.9A
12658eV
|
100x100 0
200x200 0
300x300 0 |
|
|
|
|
|
|
|
|
|
|
0.9795
12658eV |
100x100 18293
200x200 47887
300x300 71137
|
70.63
47.41
17.99 |
1.7x10^11
4.5x10^11
6.7x10^11
|
252.8--merit 726
252.8
249.8
|
|
|
|
|
|
|
|
1.0A
12658eV |
100x100 18680
200x200 47263
300x300 68925
|
18.2
46.25
67.55
|
1.2x10^11
4.2x10^11
6.2x10^11 |
244.7--merit 802
244.7
244.7
|
|
|
|
|
|
|
|
1.1
|
100x100 27195
200x200 67918
300x300 102168
|
25.18
63.35
95.39
|
2.0x10^11
5.0x10^11
7.5x10^11 |
241.0---merit 1561
|
|
|
|
|
|
|
|
1.2
|
100x100 9604
200x200 24820
300x300 38174 |
8.11
21.10
32.35
|
5.6x10^10
1.5x10^11
2.2x10^11 |
238.1---merit 645 |
|
|
|
|
|
|
|
July 28, 2008
New support for the roll has been
added. This is also a new second xtal in the mono.
We also have a new kappa arm called
skinny kappa which we want to test. We believe we are at the Se edge
and we are running in single lambda mode until motor issues are
resolved.
DJ90_1
run 1
The beam seems fine on the phospho paddle. We have the slits at
100x100 just calibrated (with a graph?).
This is 1.0 deg for 0.5 sec and it diffracts to 2.2A Rsym 0.078
0.1.66% rej.
This data collection is mostly to evaluate the behaviour of the beam
after the addition of a clamp on the roll.
This xtal also has some Semet and I would like to see some signal in
the data.
|

|
|
DJ90d4_4
Data collected at 0.9795A at X25 first graph no reject, secong graph
after reject
from the scan the infl should be at 0.9794A and peak at 0.9792A
|

|

|
|
at X29 0.4 sec for 1 deg,
collected before mounting on X25.
slits are at 100x 160 as usual. 0.9795A
diffracts to 1.9A Rsym 4.6% rej 0.22%
|

|

|
This would be more usuful if it
was at the peak and not the inf. The yellow and blue line should be
apart for low res if there is some anomalous signal.
|

|

|
June 13, 2008 (it's a
friday!)
After multi problems
with the motors mostly of the mono, there is beam in the hutch.
Motors off (pitch ,
theta and lift) in other words we are at 1.0757A and we stay there.
I am collecting on a
NAO mutant P3221 107 107 337 90 90 120.
Ring: digital feedback
on (V and H)
ah24bb4
run 4
I can't really see the beam on the phospho paddle because it's to dim.
When I open 200x200 then I can see a slight beam.
the ring current is at 198.8mA and there was a global orbit correction
right before I started collecting.
This is 0.5 deg for 10 sec and it diffracts to 2.25A Rsym 0.061
0.12% rej
the slits are 100x100.
This data collection is mostly to evaluate the behaviour of the beam
after the addition of a clamp on the roll during the shutdown.
As you can see the scaling is very quiet except for a glicth at frame
100.
|

|

|
|
|
|
|
|
|
|
|
|
|
29-30 April 2008-
with focus and defocused beam, slits 150x150
Ring: digital feedback
on (V and H)
28 April 2008-
with focus at 2.4 and slits 150x150 (because I am missing the xtal)
|
xtal_ke_5th
This
is
for 0.5 deg 3 sec exposure
no attenuator
Rsym
6.0
rejects 0.19%
The temporary clamp on the xtal-roll has been removed
killed
the previous xtal, so I had to take a small one not collected on x29.
this
scaling IS as quiet as X29
??????
The green line on the I0 plot shows a bigger drop than expected.
We did not move the roll motor after being unclamped. The pitch motor
seems to have some reliability problems.
|

|

|
xtal Ke
10ai-1 collected on april 18 2008
on X25 same xtal as below
0.5 sec with one attenuator
xtal shot at the tip
|
no I0 plot
because the data collection stopped (too short exposure time)
|
 |
|
xtal Ke
10ai-1 (Ke
10ai-1_3 collected on april 2008)
on X29
for this data collection, the scaling is pretty much typical for
this beamline.
|

|

|
|
|
|
Here
are some of the datasets collected -1 march 2008- with focus at
1.8 and slits 100x100
|
This
is
for 0.5 deg 1.25 sec
Rsym
5.6
rejects 2.2%
|
 |
 |
|
This
is
for 1 deg 2.5 sec
Rsym
5.3
rejects 2.5%
|
|

|
|
This is for 2 deg 5 sec
Rsym 6.3 rejects 2.18%
|
|
|
|
So I
adjusted the exposure time to be about equivalent per degree. That
seems to be the major factor as far as rejection.
What
really bothers me is the noise in my scaling.
Here
are
some scaling for the SAME xtal at X29
(slits v=100 H=160)
The
scalings look the same and are
very smooth. They always are.
|

X29 1deg 3 sec
Rsym 6.3 reject 3.8%
|

X29 2 deg 1 sec
Rsym6.4 reject 1.5 |
Data
collected at X25 on Feb 28, 2008 defocused at 1.8 slits 100x100
|
This
is
for 0.5 deg 10 sec
Rsym
7.1
rejects 15%
|
|
 |
|
This
is
for 1 deg 10 sec
Rsym
7.7
rejects 8%
|
|
|
|
This is for 2 deg 10 sec
Rsym 7.5 rejects 2.8%
|
|
|
|
So I
adjusted the exposure time to be about equivalent per degree. That
seems to be the major factor as far as rejection.
What
really bothers me is the noise in my scaling.
Here
are
some scaling for the SAME xtal at X29
1 deg
for
3 sec Rsym 6.3 reject 3.8%
2 deg
for
1 sec Rsym6.4 reject 1.5
the
scalings look the same and are very smooth. They always are.
|
|
|
Data
collected at X25 feb 28, 2008 focussed at 2.8 slits 100x100
|
1 deg
2
sec
Rsym
6.0
Rejects 2.2%
(6Hb_x25_3)
|
 |
 |
|
2
deg 2 sec
Rsym
7.0 Rejects 0.9%
(6Hb_x25_6)
|
 |
|
2 deg 2
sec
Rsym 7.0
Rejects 0.9%(6Hb_x25_7)
|
|
|
|
|
|
May 2008 shutdown: clamp
was removed
and, to substitute for its impact on suppressing vibrations while
allowing
freedom of motion of the second crystal assembly roll cradle, the
cradle was
re-built, i.e. it was stiffened, springs were replaced, and flex pivots
were
replaced. This succeeded in substituting properly for the removed
clamp,
in terms of eliminating vibrations, while allowing for freedom of
motion of the
cradle.
Winter 2008 cycle: the
early part of
the cycle was given to commissioning the new micro-diffractometer,
which had
its own set of issues that needed to be dealt with. Once they
were dealt
with, continued studies of the vibrations were undertaken. A
horizontal
beam instability, in the ring, was discovered, affecting not only X25
but also
X29 and other beamlines, due to the digital feedback system not being
tuned
up. As it happened, this instability, although evident in
frequency
analysis of the x-ray beam (at around 20 Hz), was too fast to cause the
~2 Hz
effect that we knew was in the beam and likely associated with the
monochromator.
At Lonny Berman’s suggestion, the second crystal assembly roll
cradle was clamped, and that appeared to eliminate this ~2 Hz nuisance
entirely, although the 20 Hz ring instability in the horizontal
direction was
still present. Test crystallography data confirmed that the ~2 Hz
nuisance was the likely cause of all the earlier problems for
crystallography
data collection, because those problems entirely disappeared upon
installation
of the clamp.
December 2007
shutdown: vibration
isolation pads were added, beneath the monochromator support, per Ed
Haas
… no impact at all on the vibrations, when beam resumed operation in
January 2008.
May 2007 shutdown: open up monochromator
chamber, and immediately “a finger was put on the problem”,
literally speaking. The second crystal assembly roll cradle, in
particular, was felt (with a finger) to be vibrating. This got
confirmed
using an accelerometer. A dampening weight was added to this
portion of
the assembly. When beam resumed operation, this showed marginal
improvement in the vibrations.
More tests and measures done during early
summer 2007: shut off all potential vibration-inducing systems at
the
beamline such as pumps, the cryogenic cooling system, even the building
air
conditioning systems in the nearby mechanical equipment room … no
impact
at all on the vibrations. Also installed added supports beneath
the monochromator
(per Mary Carlucci-Dayton’s advice) and isolate it from the upstream
portion of the beamline (by undoing a couple of clamps surrounding a
bellows)
… no impact at all on the vibrations. Added another dampening
weight to the second crystal assembly roll cradle and changed its
retaining
springs, which again led to marginal improvements in the vibrations,
but still
they were observable. Nevertheless, they were at a small enough
level
that reasonable crystallography data could be collected, and thereafter
users
were booked to collect data at X25 from July through the end of the
2007
calendar year.
Spring 2007: new optics
installed
and commissioned, vibrations (beam flickering) noticed right away,
particularly
in the horizontal direction.